ThermoFlex900Thermo Fisher
OEMThermo FisherThermo Fisher
CategoryRecirculating chillerRecirculating chiller
Wafer size——
Process node——
Introduced——
Production run——
Lifecycle——
Lifecycle milestones——
Control system——
Generation——
FamilyThermo Fisher ThermoFlex15000/20000Thermo Fisher ThermoFlex900
Cited variants—
  • ThermoFlex900/1400[1]
Specifications
Product typeThermoFlex recirculating chillers (Basic Controller)[2]—
Designed functionProvide a continuous supply of fluid at a constant temperature and volume[2]Provides a continuous supply of fluid at a constant temperature and volume[2]
Product familyThermo Scientific ThermoFlex recirculating chillers[2]—
ControllerBasic Controller[2]—
Acceptable process fluidsFiltered/single distilled water; deionized water (1-3 MΩ-cm, compensated); 0-75% ethylene glycol/water; 0-75% propylene glycol/water[2]—
Facility water connections, ThermoFlex15000-24000 inlet3/4 in cast bronze[2]—
Facility water connections, ThermoFlex15000-24000 outlet3/4 in stainless steel[2]—
Process fluid connection inletSame size as outlet on all units; stainless steel[2]—
Process fluid connection outlet, ThermoFlex7500-240001 in wrought copper[2]—
Alarm temperature setpoint range+3°C to +42°C[2]—
Setpoint range+5°C to +40°C[2]—
Serial communicationsoff, RS232, RS485[2]—
Baud rate options9600, 4800, 2400, 1200, 600, or 300 bits per second[2]—
Maintenance reminderCArE setting for preventive cleaning reminder of air and fluid filters; off, L1 (1000 hours), L2 (2000 hours), L3 (3000 hours)[2]—
Product line—ThermoFlex recirculating chillers (Basic Controller)[2]
Purpose—Designed to provide a continuous supply of fluid at a constant temperature and volume[2]
Approved fluids—Filtered/single distilled water; deionized water (1-3 MΩ-cm, compensated); 0-75% ethylene glycol/water; 0-75% propylene glycol/water[2]
Process fluid outlet connection—ThermoFlex900 - 10000: 1/2 in cast bronze[2]
Process fluid inlet connection—Same size as outlet; all units stainless steel[2]
Control options—Temperature, pressure, flow, and serial communication options are documented in the setup procedure[2]