Waferpedia

Comparison ledger

ThermoFlex15000/20000versusThermoFlex24000

Thermo Fisher ThermoFlex15000/20000, Thermo Fisher ThermoFlex24000, side by side. Every value shown is drawn from the cited encyclopedia entries.

Email
Side-by-side comparison of selected equipment models
Attribute
ThermoFlex24000Thermo Fisher
OEMThermo FisherThermo Fisher
CategoryRecirculating chillerRecirculating chiller
Wafer size——
Process node——
Introduced——
Production run——
Lifecycle——
Lifecycle milestones——
Control system——
Generation——
FamilyThermo Fisher ThermoFlex15000/20000Thermo Fisher ThermoFlex24000
Specifications
Product typeThermoFlex recirculating chillers (Basic Controller)[1]—
Designed functionProvide a continuous supply of fluid at a constant temperature and volume[1]—
Product familyThermo Scientific ThermoFlex recirculating chillers[1]—
ControllerBasic Controller[1]—
Acceptable process fluidsFiltered/single distilled water; deionized water (1-3 MΩ-cm, compensated); 0-75% ethylene glycol/water; 0-75% propylene glycol/water[1]—
Facility water connections, ThermoFlex15000-24000 inlet3/4 in cast bronze[1]—
Facility water connections, ThermoFlex15000-24000 outlet3/4 in stainless steel[1]—
Process fluid connection inletSame size as outlet on all units; stainless steel[1]—
Process fluid connection outlet, ThermoFlex7500-240001 in wrought copper[1]—
Alarm temperature setpoint range+3°C to +42°C[1]—
Setpoint range+5°C to +40°C[1]—
Serial communicationsoff, RS232, RS485[1]—
Baud rate options9600, 4800, 2400, 1200, 600, or 300 bits per second[1]—
Maintenance reminderCArE setting for preventive cleaning reminder of air and fluid filters; off, L1 (1000 hours), L2 (2000 hours), L3 (3000 hours)[1]—
Product line—ThermoFlexTM Recirculating Chillers (Basic Controller)[1]
Function—Provides a continuous supply of fluid at a constant temperature and volume[1]
Manufacturer—Thermo Fisher[2]
Equipment type—Recirculating laboratory chiller[3]
Electrical—400-460VAC, 50/60Hz, 70 Amps, 3 PH[4]
Refrigerant—R407c, 114 oz.[4]

Plan your fab

Building a Recirculating chiller process? Get a complete equipment list.

Open the fab equipment planner →
Sources (4)Every fact above is drawn from these public sources
  1. [1]nanofab.utah.edunanofab.utah.edu
  2. [2]waferpedia.comwaferpedia.com
  3. [3]questpair.comquestpair.com