API 2000 versus API 3000

SCIEX API 2000 and SCIEX API 3000, side by side. Every value shown is drawn from its cited encyclopedia entry.

Side-by-side comparison of selected equipment models
Attribute
OEMSCIEXSCIEX
CategoryPrecision OpticsPrecision Optics
Wafer size
Process node
Introduced1998
Production run1998–
Lifecycle
Lifecycle milestones1998 platform evolution reference
Control system
Generation
FamilySCIEX API 2000SCIEX API 3000
Specifications
form factorbenchtop[1]
instrument typesingle and triple quadrupole mass spectrometer products[1]
pumping platformturbo-molecular-pumped[1]
Mass Rangem/z 5-1,800[2]
Dynamic Range1 cps to 4x10^6 cps (pulse counting)[2]
Scan Speedup to 2400 amu/sec[2]
Footprint (Width x Depth x Height)104 cm x 50 cm x 64 cm[2]
Weight111 kg (244 lb)[2]
Power Requirements2 separate 207 V minimum to 242 V maximum, 15 A grounded single phase circuit[2]
Collision CellLINAC collision cell[2]LINAC collision cell[3]
manufacturerSCIEX[1]
model mention contextAPI 3000[1]
product family contextturbo-molecular-pumped single and triple quadrupole mass spectrometer products[1]
Mass AnalyzerTriple quadrupole (Q1q2Q3)[3]
Ionization SourceAtmospheric Pressure Ionization (API) with TurboIonSpray and Heated Nebulizer (APCI) options[4]
Interface TechnologyCurtain Gas interface[3]
Operating System / SoftwareWindows NT platform-based Analyst software[3]
Electrical Requirement (Main Power)207–242 VAC, 50/60 Hz, 15 A per circuit, two circuits required[4]
Power Consumption (Mass Spectrometer + Roughing Pump)220 VAC at 13–15 A (nominal)[4]
Curtain Gas RequirementUHP N2 (99.999%) at 60 psig, flow up to 3.5 L/min[4]
Gas 1 Requirement (Nebulizer/Auxiliary)Zero air or UHP N2 (99.999%) at 90 psig, flow up to 4.0 L/min[4]
Gas 2 Requirement (Heater/Nebulizer)Zero air or UHP N2 (99.999%) at 90 psig, flow up to 8.0 L/min[4]
Source Exhaust Gas RequirementClean, dry, oil-free air at 50 psig, flow up to 8.0 L/min[4]
Operating Temperature Range15–30 °C (59–86 °F)[4]
Operating Relative Humidity20–80% non-condensing[4]

Plan your fab

Building a Precision Optics process? Get a complete equipment list.

Open the fab equipment planner →
Sources (4)Every fact above is drawn from these public sources
  1. [1]en.wikipedia.orgen.wikipedia.org
  2. [2]shop.labexchange.comshop.labexchange.com
  3. [3]triadsci.comtriadsci.com
  4. [4]cdn.labwrench.comcdn.labwrench.com