Agilent

Power
To 75°C/min on a 120 V oven[3]
Vacuum
Controls all gas pressures and flows[3]
Gas delivery
Gas Chromatography[2]
Optics
Up to three detectors[3]
The Agilent 6890 GC incorporates electronic pneumatic control (EPC) for precise regulation of gas pressures and flows. Onboard sensors compensate for ambient temperature and barometric pressure changes to improve reproducibility. The system can accommodate up to two inlets and up to three detectors.[3][1]
The Agilent 6890 GC oven can be programmed from 4 to 450 °C with up to 20 ramps and 21 plateaus. Fast ramp rates up to 75 °C per minute are possible. The oven can cool from 300 °C to 50 °C in 4.5 minutes. Electronic pneumatic control regulates carrier and detector gases to 0.001 psi and 0.05 mL/min.[3][1][5]
The 6890 GC supports detectors including flame ionization detector (FID), thermal conductivity detector (TCD), micro-electron capture detector (μ-ECD), nitrogen-phosphorus detector (NPD), and flame photometric detector (FPD). The system can be configured with up to three detectors simultaneously, although typical configurations carry one or two.[3][1]
Official Agilent support for the 6890N ended on July 31, 2015.
Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.
Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.
No research found yet — worked with this tool? Share what you know.
The oven can be programmed from 4 °C to 450 °C.[3]
The instrument can accommodate up to two inlets.[3]
Supported detectors include flame ionization detector (FID), thermal conductivity detector (TCD), micro-electron capture detector (μ-ECD), nitrogen-phosphorus detector (NPD), and flame photometric detector (FPD).[3][1]
Electronic pneumatic control regulates gas pressures to 0.001 psi and flows to 0.05 mL/min.[3]
Official Agilent support for the 6890N ended on July 31, 2015.[1]
The following facts about the 6890 GC are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.
No publicly documented variants, configuration options, or revision breakpoints of the 6890 GC are on record.
Answerable by: an OEM product catalog or an engineer who ordered or specified the tool
The control-system platform and OS era of the 6890 GC are not on record.
Answerable by: an engineer who operated it or OEM installation records
No publicly documented failure modes or field errata for the 6890 GC are on record.
Answerable by: a field service engineer, process engineer, or maintenance technician
The process node or technology generation of the 6890 GC is not on record.
Answerable by: an OEM datasheet or a fab qualification report
No publicly documented compatible parts, consumables, or accessories for the 6890 GC are on record.
Answerable by: an OEM parts catalog or a service engineer
No research found yet — worked with this tool? Share what you know.
Last updated Oct 12, 2026.