The Tektronix 7854 is a 400 MHz waveform processing oscilloscope that combines analog and digital storage capability. The 7854 features a 10-bit digitizer, programmable keystroke operation, and IEEE-488 (GPIB) interface.[1][2]

The Tektronix 7854 oscilloscope was introduced in 1980. The 7854 is a combined analog and digital waveform-processing oscilloscope. The 7854 offers a bandwidth of 400 MHz in analog mode. The 7854 can store repetitive waveforms up to 400 MHz. The 7854 mainframe accepts two vertical and two horizontal plug-in modules from the 7000 series.[2][1]
The 7854 uses a digitizer that simultaneously samples X and Y signal values at a rate of 3.5 µs per point. The digitizer employs sample-and-hold circuits based on Schottky diode bridges and a single 10-bit successive-approximation ADC. The digitized Y value is written into memory via direct memory access (DMA) bypassing the CPU. The CPU is a Texas Instruments TMS9900 16-bit microprocessor with 32+6 KByte ROM and 8 KByte RAM. The 7854 can store repetitive waveforms up to 400 MHz, and up to 14 GHz with the 7811/7S12 sampling plug-ins. The fastest calibrated sweep rate is 500 ps/div.[2][1]
The 7854 includes a standard IEEE-488 (GPIB) interface for remote control and automated test setups. The GPIB allows waveform data, constants, keystroke programs, and text to be transmitted or received. The 7854 can process waveform data and send answers over the GPIB instead of raw data, reducing transfer time.[1]
The 7854 can perform time-domain reflectometry (TDR) when used with appropriate sampling plug-ins. The oscilloscope supports signal averaging to uncover signals buried in noise. Cursor measurements allow amplitude and time measurements on waveforms. The waveform calculator can perform arithmetic operations such as multiplication of voltage and current to yield power.[4][1]
| Designation | Generation | Vintage | Changes | Source |
|---|---|---|---|---|
| Option 02 | — | — | X-Y Phase Correction | w140.com[1] |
| Option 03 | — | — | EMC Modification | w140.com[1] |
| Option 78 | — | — | P11 Phosphor CRT | w140.com[1] |
| Option 2D | — | — | 4K Expanded memory storage for up to 40 waveforms and 2000 keystrokes | w140.com[2] |
| Option 0D | — | — | Delete GPIB and remote Waveform Calculator, one waveform storage | w140.com[1] |
Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.
Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.
No research found yet — worked with this tool? Share what you know.
The digitizer has 10-bit resolution, with a vertical resolution of 0.01 division.[1]
Options include Option 02 (X-Y phase correction), Option 03 (EMC modification), Option 78 (P11 phosphor), and Option 2D (expanded memory for 40 waveforms and 2000 keystrokes).[2]
The following facts about the 7854 Oscilloscope are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.
The control-system platform and OS era of the 7854 Oscilloscope are not on record.
Answerable by: an engineer who operated it or OEM installation records
No publicly documented failure modes or field errata for the 7854 Oscilloscope are on record.
Answerable by: a field service engineer, process engineer, or maintenance technician
The process node or technology generation of the 7854 Oscilloscope is not on record.
Answerable by: an OEM datasheet or a fab qualification report
No publicly documented compatible parts, consumables, or accessories for the 7854 Oscilloscope are on record.
Answerable by: an OEM parts catalog or a service engineer
No publicly hosted manuals, SOPs, or datasheets for the 7854 Oscilloscope are on record.
Answerable by: university cleanroom staff or an OEM application specialist
No research found yet — worked with this tool? Share what you know.
Last updated Sep 27, 2026.