Waferpedia

Varian

2300C/D

Produced 1987–2007Varian 2300C/D family
Research Quality: 50% complete

The Varian 2300C/D is a variant of the Clinac 2100C/2100CD linear accelerator used for radiation therapy. The Varian 2300C/D was manufactured from 1987 to 2007. The Varian 2300C/D uses a klystron power source.[1]

Varian2300C/D
No photo available yet

What it is

The Varian 2300C/D is a medical linear accelerator used for external-beam radiation therapy. The 2300C/D belongs to the Varian Clinac 2100C/2100CD generation of linear accelerators. The Clinac 2100C/2100CD generation was manufactured from 1987 through 2007. The machine uses a klystron as its microwave power source. The generation offers photon energy configurations of 6 MV and 10/15/18/25 MV. The generation also provides electron beam capability. A Silhouette option for smaller treatment-room layouts was available for the family.[1]

How it works

The Clinac 2100C/2100CD generation, which includes the 2300C/D, offers photon beam configurations of 6 MV and 10/15/18/25 MV. The maximum photon dose rate of the generation is 600 MU/min. The system provides electron beam therapy at a maximum electron dose rate of 1000 MU/min. Beam shaping options include an 80-leaf multi-leaf collimator for 40 x 40 cm fields and a 120-leaf multi-leaf collimator. In the 120-leaf version, the central 20 cm of the field uses 5 mm leaves, and the outer 20 cm of the field uses 10 mm leaves. Electronic portal imaging options for the generation include LC250, aS500, and aS1000 devices, available as optional equipment. Couch options for the generation include Exact ETR, with the Exact couch available as an optional configuration.[1]

Applications

The Clinac 2100C/2100CD generation, which includes the 2300C/D, supports three-dimensional conformal radiation therapy and intensity-modulated radiation therapy. Stereotactic radiosurgery and stereotactic body radiation therapy are optional capabilities for the generation. Stereotactic delivery uses optional cones or a mini multi-leaf collimator. Image-guided radiation therapy is possible with the optional On-Board Imager kilovoltage imaging system. Cone-beam computed tomography is available only when the On-Board Imager is fitted. Volumetric-modulated arc therapy is also available only when the On-Board Imager is fitted.[1]

  • 3D conformal therapy
  • IMRT
  • SRS/SBRT (optional)

What do the numbers mean?

Dimensions & facilities1

Multi-leaf collimator (MLC)
80 leaves (field size 40x40 cm); 120 leaves available (field size 40x40 cm, central 20 cm 5 mm leaf width, outer 20 cm 10 mm leaf width)[1]
Accurate?

Configuration & options14

Power source
Klystron[1]
Accurate?
Photon energy configuration
6 & 10/15/18/25 MV[1]
Accurate?
Maximum dose rate (photons)
600 MU/min[1]
Accurate?
Electron energies
Yes[1]
Accurate?
Maximum dose rate (electrons)
1000 MU/min[1]
Accurate?
Portal imager (EPID)
LC250, aS500, aS1000 (optional)[1]
Accurate?
kV imaging for IGRT
OBI (optional)[1]
Accurate?
Cone-beam CT
Only with OBI[1]
Accurate?
Treatment delivery
3D, IMRT, SRS/SBRT (optional)[1]
Accurate?
Stereotactic radiosurgery
Optional cones or mMLC[1]
Accurate?
Treatment couch
Exact ETR, Exact (optional)[1]
Accurate?
iCBCT
No[1]
Accurate?
ART (adaptive radiotherapy)
No[1]
Accurate?
VMAT
Only with OBI[1]
Accurate?

Vintage & configurations

  1. 1987Production start[1]

    Start of manufacture for the Clinac 2100C/2100CD family, which includes the 2300C/D variant

  2. 2007Production end[1]

    End of manufacture for the Clinac 2100C/2100CD family, which includes the 2300C/D variant

Interested in this tool?
Need help with this tool?Embed spec card

What replaced it?

  • siblingVarian 23EX— Another variation of the Clinac 2100C/2100CD line listed alongside 2300C/Dsource[1]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What photon energies are available on the 2300C/D?

The Clinac 2100C/2100CD generation, which includes the 2300C/D, offers photon energy configurations of 6 MV and 10/15/18/25 MV.[1]

What is the maximum photon dose rate?

The maximum photon dose rate of the generation is 600 MU/min.[1]

Does the machine deliver electron beams?

The 2300C/D family provides electron beam therapy. The maximum electron dose rate is 1000 MU/min.[1]

Which multi-leaf collimator options are offered?

The generation offers an 80-leaf multi-leaf collimator for 40 x 40 cm fields and a 120-leaf multi-leaf collimator. The 120-leaf design uses 5 mm leaves across the central 20 cm of the field and 10 mm leaves across the outer 20 cm.[1]

Can the machine perform stereotactic treatments?

Stereotactic radiosurgery and stereotactic body radiation therapy are optional capabilities for the generation. Stereotactic delivery uses optional cones or a mini multi-leaf collimator.[1]

What imaging systems are available?

The generation offers optional electronic portal imaging with LC250, aS500, or aS1000 devices. A kilovoltage On-Board Imager is available as an optional system for image-guided radiation therapy. Cone-beam computed tomography and volumetric-modulated arc therapy are available only when the On-Board Imager is fitted.[1]

Not publicly documented

The following facts about the 2300C/D are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented variants, configuration options, or revision breakpoints of the 2300C/D are on record.

    Answerable by: an OEM product catalog or an engineer who ordered or specified the tool

  • The control-system platform and OS era of the 2300C/D are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the 2300C/D are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the 2300C/D is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

  • No publicly documented compatible parts, consumables, or accessories for the 2300C/D are on record.

    Answerable by: an OEM parts catalog or a service engineer

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (1)Every fact above is drawn from these public sources
  1. [1]oncologysystems.com — oncologysystems.comoncologysystems.com
Was this page accurate?

Last updated Sep 22, 2026.