Waferpedia

Varian

640 IR FTIR

Precision Optics
Research Quality: 40% complete

Provisional entry — research in progress

This page was generated automatically from cited public sources and hasn't completed the full research pass yet. Every specification shown carries its source; more detail is added as research completes.

The Varian 640-IR is a Fourier Transform Infrared (FT-IR) spectrometer. The Varian 640-IR is based on a 38 mm dynamically aligned, 60° mechanical bearing Michelson interferometer. The Varian 640-IR is available in mid-IR or near-IR configurations.[1]

Varian640 IR FTIR
No photo available yet

What it is

The Varian 640 IR is a benchtop spectrometer manufactured by Varian, Inc. The instrument uses a 38 mm dynamically aligned, 60° mechanical bearing Michelson interferometer. The Varian 640 IR is available in mid-IR and near-IR configurations. The standard mid-IR range is 7,900 to 375 cm⁻¹ and the standard near-IR range is 11,000 to 2,000 cm⁻¹. The instrument achieves an unapodized optical resolution of less than 0.18 cm⁻¹. The spectrometer enclosure is sealed and desiccated, with a tropical (moisture-resistant) option available. The Varian 640 IR uses a 24-bit delta-sigma A/D converter and communicates via USB 2. The instrument weighs 80 kg (176 pounds) and measures 70.8 x 75.6 x 34.4 cm (27.9 x 29.8 x 13.5 in).[3][1][2]

How it works

The Varian 640 IR uses a 38 mm dynamically aligned, 60° mechanical bearing Michelson interferometer. The interferometer design provides stability and reproducibility. The instrument includes an air-cooled ceramic source for mid-IR and a tungsten-halogen source for near-IR. The standard detector is a cooled DLaTGS for mid-IR and a PbSe detector for near-IR. The interferometer is dynamically aligned to maintain optical alignment over time.[1]

Applications

The Varian 640 IR is suited to QA/QC applications, identification and verification analyses, and is an ideal educational tool. The instrument is compatible with a wide range of accessories, including single-point microscopy, microscopy imaging, macro imaging, patented ATR imaging, and hyphenated techniques such as thermogravimetric analysis (TGA-IR).[1]

What do the numbers mean?

Optics & imaging7

Spectral resolution (unapodized)
< 0.18 cm-1 (Mid IR), < 0.25 cm-1 (Near IR)[1]
Accurate?
Signal-to-noise ratio
6,000:1 peak-to-peak (absorbance = 7.2 x 10-5) measured at 5 seconds, 4 cm-1 resolution[1]
Accurate?
Mid IR detector
Cooled DLaTGS[1]
Accurate?
Near IR detector
PbSe[1]
Accurate?
Wavelength range
11000 to 735 cm-1[2]
Accurate?
Wavelength accuracy
>0.01 cm-1[2]
Accurate?
Wavelength precision
>0.005 cm-1[2]
Accurate?

Control & software1

Spectrometer interface
USB 2[1]
Accurate?

Dimensions & facilities3

Sample compartment dimensions (W x D x H)
23.2 x 27.6 x 15.4 cm (9.1 x 10.9 x 6.1 in)[1]
Accurate?
Spectrometer dimensions (W x D x H)
70.8 x 75.6 x 34.4 cm (27.9 x 29.8 x 13.5 in)[1]
Accurate?
Weight
80 kg (176 pounds)[1]
Accurate?

Configuration & options11

Interferometer type
38 mm dynamically aligned, 60° mechanical bearing Michelson[1]
Accurate?
Spectral range (Mid IR)
7,900–375 cm-1[1]
Accurate?
Spectral range (Near IR)
11,000 - 2,000 cm-1[1]
Accurate?
Spectrometer enclosure
Sealed and dessicated (Standard); Tropical moisture-resistant windows (Optional)[1]
Accurate?
A/D converter dynamic range
Delta-Sigma (24 bits)[1]
Accurate?
External ports
3 (left, right and rear emission)[1]
Accurate?
Mid IR source
Ceramic[1]
Accurate?
Mid IR beamsplitter
Optimized KBr[1]
Accurate?
Near IR source
Tungsten-halogen[1]
Accurate?
Near IR beamsplitter
NIR quartz[1]
Accurate?
Scan speed
>40 scans per second[2]
Accurate?
Interested in this tool?
Embed spec card

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Not publicly documented

The following facts about the 640 IR FTIR are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the 640 IR FTIR are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • No publicly documented variants, configuration options, or revision breakpoints of the 640 IR FTIR are on record.

    Answerable by: an OEM product catalog or an engineer who ordered or specified the tool

  • The control-system platform and OS era of the 640 IR FTIR are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the 640 IR FTIR are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the 640 IR FTIR is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (3)Every fact above is drawn from these public sources
  1. [1]manuallib.commanuallib.commanuallib.com
  2. [2]americanlaboratorytrading.comamericanlaboratorytrading.comamericanlaboratorytrading.com
  3. [3]Equipment inventory listing
Was this page accurate?

Last updated Aug 24, 2026.

Compare with similar tools

View all →