Tektronix 485 Oscilloscope family
Research Quality: 30% complete
Email

The Tektronix 485 is a portable, dual-channel wide-band oscilloscope. The 485 provides DC to 350 MHz bandwidth with 50 Ω input impedance and DC to 250 MHz bandwidth with 1 MΩ input impedance. The 485 has a calibrated sweep rate from 1 ns/div to 0.5 s/div.[1]

485 Oscilloscope — Inventory photo
Fig. 01485 OscilloscopeInventory photo[2]

What is it?

The Tektronix 485 is a portable, dual-channel wide-band oscilloscope with selectable input impedance, providing up to 350 MHz bandwidth in 50 Ω mode and 250 MHz in 1 MΩ mode.

What do the numbers mean?

Power & electrical10

Frequency
48–440 Hz (50–400 Hz Maximum)[2]
Accurate?
Watts Max
80[2]
Accurate?
Amps Max
0.9 A 115 V[2]
Accurate?
Bandwidth (50 Ω)
DC to at least 350 MHz[1]
Accurate?
Bandwidth (1 MΩ)
DC to at least 250 MHz[1]
Accurate?
Deflection factor
5 mV/div to 5 V/div in 10 calibrated steps, variable to 12.5 V/div uncalibrated[1]
Accurate?
Calibrator frequency
Selectable 1 kHz and 1 MHz[1]
Accurate?
Calibrator amplitude
5 V into 1 MΩ load, 0.5 V into 50 Ω load[1]
Accurate?
Line voltage range
115 V (90–136 V) or 230 V (180–272 V); 48–440 Hz[1]
Accurate?
Power consumption
60 W maximum at 115 V, 60 Hz; 0.9 A max[1]
Accurate?

Control & software1

CRT display area
8 horizontal x 10 vertical divisions (0.8 cm/div)[1]
Accurate?

Dimensions & facilities3

Dimensions (H x W x D)
6 9/16 in x 12 in (with handle) x 18 1/2 in (with panel cover)[1]
Accurate?
Net weight (without accessories)
21 lb[1]
Accurate?
Net weight (with panel cover and accessories)
24 lb[1]
Accurate?

Configuration & options9

Line Fuse
3 A Fast[2]
Accurate?
Risetime (50 Ω)
1 ns[1]
Accurate?
Risetime (1 MΩ)
1.4 ns[1]
Accurate?
Sweep rate range
1 ns/div to 0.5 s/div in 27 calibrated steps[1]
Accurate?
Input impedance (50 Ω)
50 Ω within 0.5%[1]
Accurate?
Input impedance (1 MΩ)
1 MΩ within 1% paralleled by approximately 20 pF[1]
Accurate?
Accelerating potential
21 kV[1]
Accurate?
Operating temperature
−15°C to +55°C[1]
Accurate?
Storage temperature
−55°C to +75°C[1]
Accurate?

Vintage & configurations

Documented models & variants

DesignationGenerationVintageChangesSource
485-1——No external trigger viewarchive.org[1]
485-2——No external trigger view; only 50 Ω vertical input impedancearchive.org[1]
R485——Rackmount version of the 485 oscilloscopearchive.org[1]
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Frequency48–440 Hz (50–400 Hz Maximum)[2]
  • Watts Max80[2]
  • Amps Max0.9 A 115 V[2]
  • Bandwidth (50 Ω)DC to at least 350 MHz[1]
  • Bandwidth (1 MΩ)DC to at least 250 MHz[1]
  • Deflection factor5 mV/div to 5 V/div in 10 calibrated steps, variable to 12.5 V/div uncalibrated[1]
  • Calibrator frequencySelectable 1 kHz and 1 MHz[1]
  • Calibrator amplitude5 V into 1 MΩ load, 0.5 V into 50 Ω load[1]
  • Line voltage range115 V (90–136 V) or 230 V (180–272 V); 48–440 Hz[1]
  • Power consumption60 W maximum at 115 V, 60 Hz; 0.9 A max[1]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Not publicly documented

The following facts about the 485 Oscilloscope are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the 485 Oscilloscope are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • The control-system platform and OS era of the 485 Oscilloscope are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the 485 Oscilloscope are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the 485 Oscilloscope is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

  • No publicly documented compatible parts, consumables, or accessories for the 485 Oscilloscope are on record.

    Answerable by: an OEM parts catalog or a service engineer