Keysight4142 B
No photo available yet

What is it?

The Keysight / Agilent 4142B is a modular semiconductor parameter tester mainframe with eight slots for plug-in measurement units.

What do the numbers mean?

Power & electrical12

Integrated measurement functions
Current/voltage sources, nanovoltmeters, and picoammeters in one instrument[1]
Accurate?
Source Monitor Unit (SMU) sensitivity
Forces voltage while measuring currents down to 20 fA; forces current while monitoring voltages down to 40 µV[1]
Accurate?
HP 41420A High Power SMU
2 slots; 40 µV to 200 V, 20 fA to 1 A (±200 V and ±1 A)[1]
Accurate?
HP 41421B Medium Power SMU
1 slot; 40 µV to 100 V, 20 fA to 100 mA (±100 V and ±100 mA)[1]
Accurate?
HP 41422A High Current Unit (HCU)
2 slots; pulse mode only; 100 µs pulsed measurements up to ±10 V and 10 A[1]
Accurate?
HP 41423A High Voltage Unit (HVU)
2 slots; DC or pulsed measurements to ±1000 V; low-leakage design measures currents down to 2 pA[1]
Accurate?
HP 41425A Analog Feedback Unit (AFU)
1 slot; used with two SMUs for analog voltage/current searches; finds target values in 12 ms[1]
Accurate?
Maximum configuration coverage
2000 V max using two HVUs; 20 A (pulsed) using two HCUs[1]
Accurate?
Maximum common-to-ground voltage
±42 V[1]
Accurate?
Ground Unit (GNDU)
High-current SMU set to 0 V, used as measurement ground; Kelvin connection; maximum offset voltage ±500 µV; current range ±1.6 A[1]
Accurate?
Power requirements
100/120/220 V ±10%; 240 V -10%/+5%; maximum 750 VA (48-66 Hz)[1]
Accurate?
Typical measurement time (SMU)
HP 41420A/HP 41421B SMU (20 V/100 mA range, spot measurement): force I or V in 3.5 ms; measure I or V in 4.0 ms[1]
Accurate?

Control & software1

Interfaces
HP-IB; external trigger input (TTL, negative logic, minimum pulse width 100 µs); external trigger output (TTL, negative logic)[1]
Accurate?

Dimensions & facilities2

Dimensions
426 mm W × 235 mm H × 676 mm D[1]
Accurate?
Weight
Mainframe approx. 23 kg; plug-in units 2-3 kg each[1]
Accurate?

Configuration & options9

Architecture
Mainframe with 8 slots for plug-in measurement units[1]
Accurate?
Maximum output power
Sum of all measurement units' instantaneous output power cannot exceed 32 W[1]
Accurate?
Program memory
Approx. 2000 program steps, groupable into 99 subroutines[1]
Accurate?
Data memory
Maximum 4095 data values (binary) or 1023 data values (ASCII)[1]
Accurate?
Measurement modes
Plug-in units operate individually or in synchronized modes to perform ten different types of measurements[1]
Accurate?
Self-test and calibration
Self-test at power-up; offset errors in each measurement unit auto-calibrated every 30 minutes[1]
Accurate?
Operating temperature
5 °C to 40 °C[1]
Accurate?
Operating humidity
5% to 80% RH[1]
Accurate?
Storage temperature and humidity
-40 °C to 65 °C; <90% RH at 65 °C[1]
Accurate?
Interested in this tool?
Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Integrated measurement functionsCurrent/voltage sources, nanovoltmeters, and picoammeters in one instrument[1]
  • Source Monitor Unit (SMU) sensitivityForces voltage while measuring currents down to 20 fA; forces current while monitoring voltages down to 40 µV[1]
  • HP 41420A High Power SMU2 slots; 40 µV to 200 V, 20 fA to 1 A (±200 V and ±1 A)[1]
  • HP 41421B Medium Power SMU1 slot; 40 µV to 100 V, 20 fA to 100 mA (±100 V and ±100 mA)[1]
  • HP 41422A High Current Unit (HCU)2 slots; pulse mode only; 100 µs pulsed measurements up to ±10 V and 10 A[1]
  • HP 41423A High Voltage Unit (HVU)2 slots; DC or pulsed measurements to ±1000 V; low-leakage design measures currents down to 2 pA[1]
  • HP 41425A Analog Feedback Unit (AFU)1 slot; used with two SMUs for analog voltage/current searches; finds target values in 12 ms[1]
  • Maximum configuration coverage2000 V max using two HVUs; 20 A (pulsed) using two HCUs[1]
  • Maximum common-to-ground voltage±42 V[1]
  • Ground Unit (GNDU)High-current SMU set to 0 V, used as measurement ground; Kelvin connection; maximum offset voltage ±500 µV; current range ±1.6 A[1]
  • Power requirements100/120/220 V ±10%; 240 V -10%/+5%; maximum 750 VA (48-66 Hz)[1]
  • Typical measurement time (SMU)HP 41420A/HP 41421B SMU (20 V/100 mA range, spot measurement): force I or V in 3.5 ms; measure I or V in 4.0 ms[1]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Not publicly documented

The following facts about the 4142 B are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the 4142 B are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • No publicly documented variants, configuration options, or revision breakpoints of the 4142 B are on record.

    Answerable by: an OEM product catalog or an engineer who ordered or specified the tool

  • The control-system platform and OS era of the 4142 B are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the 4142 B are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the 4142 B is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

No research found yet — worked with this tool? Share what you know.