S 4100 FE-SEM

Email

The Hitachi S 4100 FE-SEM is a field emission scanning electron microscope. The Hitachi S 4100 FE-SEM has manually operated sample handling. The Hitachi S 4100 FE-SEM uses a cold field emission source.[1][2]

S 4100 FE-SEM — Inventory photo
Fig. 01S 4100 FE-SEMInventory photo[3]

Power

Single phase AC 220 or 240 volt, 50/60 Hz[3]

Vacuum

prepumped[3]

Optics

2.0nm at 30 KV[3]

What it is

The Hitachi S 4100 FE-SEM is a field emission scanning electron microscope. The Hitachi S 4100 FE-SEM is used in metrology and inspection. The Hitachi S 4100 FE-SEM has a cold field emission source. The Hitachi S 4100 FE-SEM uses manually operated sample handling, including a manually operated 5 axis specimen chamber and a manually operated specimen exchange port.[3][1][2]

How it works

The Hitachi S 4100 FE-SEM uses a cold field emission electron gun. The Hitachi S 4100 FE-SEM uses electromagnetic lenses, an octopole electromagnetic stigmator, and a 2 stage electromagnetic scanning coil. The Hitachi S 4100 FE-SEM is specified for secondary electron image resolution of 2.0 nm at 30 KV and magnification from 20X to 300KX. The Hitachi S 4100 FE-SEM has a 4 position externally selectable objective aperture.[3][1]

The Hitachi S 4100 FE-SEM has a prepumped airlock. The Hitachi S 4100 FE-SEM has full automatic vacuum operation with pneumatic valve control. The Hitachi S 4100 FE-SEM is specified for ultimate vacuum of 10(-7) Pa in the electron gun chamber and 10(-4) Pa in the specimen chamber. The Hitachi S 4100 FE-SEM uses ion pumps rated at 60 L/s and 20 L/s, and the specimen chamber uses a DP rated at 570 L/s.[3][1]

Where it fits in the process flow

General reference — not yet source-verified

A metrology and inspection SEM is typically placed in analytical and defect review workflows rather than in the main material deposition or pattern transfer path. A tool of this class is commonly used after patterning, etch, film formation, cleaning, or planarization steps when engineers need direct imaging of surface features or defects. A tool of this class also supports failure analysis and process characterization by providing higher detail than routine optical inspection.

Applications

The Hitachi S 4100 FE-SEM supports secondary electron imaging and offers scanning modes that include normal, reduced area, line scan, photo scan, spot position, and split screen. The Hitachi S 4100 FE-SEM uses dual 12 inch monitors for image display. The Hitachi S 4100 FE-SEM includes real time processing, auto brightness and contrast control, dynamic stigmator autofocus, frame averaging, frame integration, and contrast conversion. The Hitachi S 4100 FE-SEM can be upgraded with a National Instruments Vision Acquisition computer imaging system, and optional accessories include EDX at 30 degree take off angle and digital image capturing with Orion-6 software.[3]

What do the numbers mean?

Power & electrical1

Installation power
Single phase AC 220 or 240 volt, 50/60 Hz[3]
Accurate?

Vacuum & pumping8

Airlock
prepumped[3]
Accurate?
Vacuum system operation
Full automatic operation with pneumatic valve control[3]
Accurate?
Ultimate vacuum, electron gun chamber
10(-7) Pa[3]
Accurate?
Ultimate vacuum, specimen chamber
10(-4) Pa[3]
Accurate?
Ion pump
60 L/s X1, 20 L/s X2[3]
Accurate?
Specimen chamber pump
DP (570 L/s)[3]
Accurate?
Foreline
Rotary pump X2[3]
Accurate?
Ultimate vacuum
10(-7) Pa in electron gun chamber, 10(-4) Pa in specimen chamber[3]
Accurate?

Optics & imaging5

Secondary electron image resolution
2.0nm at 30 KV[3]
Accurate?
Magnification
20X - 300KX[3]
Accurate?
Electron Optics:[3]
Accurate?
Lens type
electromagnetic[3]
Accurate?
Objective aperture
4 position externally selectable[3]
Accurate?

Control & software3

Can be upgraded with a National Instruments Vision Acquisition computer imaging system[3]
Accurate?
Optional accessory
Digital image capturing: Orion-6 software[3]
Accurate?
Digital image capturing
Orion-6 software, optional[3]
Accurate?

Configuration & options32

Manually operated 5 axis specimen chamber[3]
Accurate?
Manually operated specimen exchange port[3]
Accurate?
Specifications:[3]
Accurate?
Performance[3]
Accurate?
Electron gun
cold field emission source[3]
Accurate?
Stigmator
octopole electromagnetic[3]
Accurate?
Scanning coil
2-stage electromagnetic[3]
Accurate?
Sample Chamber:[3]
Accurate?
Size Type I[3]
Accurate?
Max sample size 100mm diameter[3]
Accurate?
Stage motion
5 axis manual, X/Y: 25mm, Z: 5-30mm, T: -5° - +45°, R: 360° (continuous)[3]
Accurate?
Draw-out door
Max sample size: 150mm diameter[3]
Accurate?
Size Type II[3]
Accurate?
Instrument type
Field emission scanning electron microscope[1]
Accurate?
Sample handling
Manually operated sample handling[1]
Accurate?
Manually operated 5 axis specimen chamber
Present[3]
Accurate?
Manually operated specimen exchange port
Present[3]
Accurate?
Maximum sample size, Size Type I
100mm diameter[3]
Accurate?
Stage motion, Size Type I
5 axis manual, X/Y: 25mm, Z: 5-30mm, T: -5° - +45°, R: 360° (continuous)[3]
Accurate?
Draw-out door maximum sample size
150mm diameter[3]
Accurate?
Maximum sample size, Size Type II
150mm diameter[3]
Accurate?
Stage motion, Size Type II
5 axis manual, X/Y: 100mm/50mm, Z: 5 - 35mm, T: -5° - +60°, R: 360° (continuous)[3]
Accurate?
Image display
Dual 12" monitors[3]
Accurate?
Scanning mode
Normal, reduced area, line scan, photo scan, spot position, split screen[3]
Accurate?
Scanning speed
TV, 0.3, 2, 9, 25, 35, 100, 160, 320 s/frame[3]
Accurate?
Optional accessory
EDX at 30° take off angle[3]
Accurate?
Optional accessory
Chilled water circulator: 10-20° C, 1.0-1.5 L/m for DP only[3]
Accurate?
Grounding
Independent grounding 100 Ohms or less[3]
Accurate?
Vintage
1993[3]
Accurate?
Stage motion
5 axis manual, X/Y: 100mm/50mm, Z: 5 - 35mm, T: -5° - +60°, R: 360° (continuous)[3]
Accurate?
Specimen chamber
DP (570 L/s)[3]
Accurate?
EDX
at 30° take off angle[3]
Accurate?
Interested in this tool?
Need help with this tool?Embed spec card

What does it need to run?

Site utility requirements, footprint, and infrastructure needed to install and operate this tool. Sourced from public records.

  • Airlockprepumped[3]
  • Vacuum system operationFull automatic operation with pneumatic valve control[3]
  • Ultimate vacuum, electron gun chamber10(-7) Pa[3]
  • Ultimate vacuum, specimen chamber10(-4) Pa[3]
  • Ion pump60 L/s X1, 20 L/s X2[3]
  • Specimen chamber pumpDP (570 L/s)[3]
  • ForelineRotary pump X2[3]
  • Installation powerSingle phase AC 220 or 240 volt, 50/60 Hz[3]
  • Ultimate vacuum10(-7) Pa in electron gun chamber, 10(-4) Pa in specimen chamber[3]

Where are the manuals?

Generated from public-source data on file. Enter your email to access — nothing is published; details are routed privately.

Not publicly documented

Field notes

No research found yet — worked with this tool? Share what you know.

Worked with this tool? Help preserve its knowledge →

Frequently asked questions

What type of electron source does the Hitachi S 4100 FE-SEM use?

The Hitachi S 4100 FE-SEM uses a cold field emission source.[3][1][2]

What resolution and magnification are specified for the Hitachi S 4100 FE-SEM?

The Hitachi S 4100 FE-SEM is specified for secondary electron image resolution of 2.0 nm at 30 KV and magnification from 20X to 300KX.[3][1]

How is specimen handling configured on the Hitachi S 4100 FE-SEM?

The Hitachi S 4100 FE-SEM has a manually operated 5 axis specimen chamber and a manually operated specimen exchange port.[3][1]

What vacuum system details are specified for the Hitachi S 4100 FE-SEM?

The Hitachi S 4100 FE-SEM has full automatic vacuum operation with pneumatic valve control. The Hitachi S 4100 FE-SEM is specified for ultimate vacuum of 10(-7) Pa in the electron gun chamber and 10(-4) Pa in the specimen chamber. The Hitachi S 4100 FE-SEM uses ion pumps rated at 60 L/s and 20 L/s, and the specimen chamber uses a DP rated at 570 L/s.[3]

Not publicly documented

The following facts about the S 4100 FE-SEM are absent from this record as of this revision. First-hand knowledge or a citation closes a gap; every submission is reviewed before publication.

  • No publicly documented production dates or lifecycle milestones (introduction, end of production, EOL) for the S 4100 FE-SEM are on record.

    Answerable by: OEM historical records or a trade-press announcement

  • No publicly documented variants, configuration options, or revision breakpoints of the S 4100 FE-SEM are on record.

    Answerable by: an OEM product catalog or an engineer who ordered or specified the tool

  • The control-system platform and OS era of the S 4100 FE-SEM are not on record.

    Answerable by: an engineer who operated it or OEM installation records

  • No publicly documented failure modes or field errata for the S 4100 FE-SEM are on record.

    Answerable by: a field service engineer, process engineer, or maintenance technician

  • The process node or technology generation of the S 4100 FE-SEM is not on record.

    Answerable by: an OEM datasheet or a fab qualification report

No research found yet — worked with this tool? Share what you know.

Sources & citations

Sources (3)Every fact above is drawn from these public sources
  1. [1]waferpedia.com — waferpedia.comwaferpedia.com
  2. [2]Hitachi S 4100 FE-SEM family | Waferpedia — waferpedia.comwaferpedia.com
  3. [3]Inventory photo
Was this page accurate?

Last updated Oct 16, 2026.

Compare with similar tools

View all →